Real-time PCR System GT-96 is an instrument and equipment based on quantitative fluorescence PCR technology. It integrates PCR amplification, detection and data analysis into one, which can provide fast, accurate and flexible detection services. Its unique constant current power supply and temperature control mode fully guarantee the system performance. The self-developed real-time PCR system can customize different software functions according to user requirements, equipped with comprehensive analysis functions and custom report formats, which fully meets customers' diverse needs for data export formats in different application scenarios.

 

Model4 Channel /6 Channel


  • SKU: 0202068/0202069
  • Category: Real-time PCR System
  • Tags:
      Geneture,Real-time PCR System,GT-96,Fluorescence,qPCR
  1. Precision gradient: unique six-block design allows simultaneous or independent runs with temperature, time and volume gradients.

  2. Accurate thermal control: thermo-compensation technology maintains ±0.1 °C uniformity across all wells.

  3. Sensitive optics: maintenance-free, long-life LED source; upgraded fiber-optic transmission reduces fluorescence loss for reliable detection.

  4. Smart analysis: high-resolution melt and amplification curve analysis with intuitive interface and one-touch start.

  5. Broad application: qualitative, absolute quantification, relative quantification, multiplex quantification, genotyping, standard melt curve, SNP, HRM, and isothermal amplification modes.

  6. Open platform: reagent- and consumable-agnostic, supporting complex assay portfolios.

Additional Information

Performance

Parameter

Dimension (W×D×H)

450 mm × 340 mm × 240 mm

Weight

17 KG

Voltage / Frequency / Power

AC 100 ~ 240V, 50Hz, 800 W

Interface

2 × USB

Operation Temperature

15 ~ 30 °C

Operation Humidity

0% ~ 85%

Transportation and Storage Temperature

-20 ~ 55 °C

Transportation and Storage Relative Humidity

0% ~ 85%

Applicable consumables

0.2 mL PCR 8-Tube strips, 0.2 mL PCR tube with cap,96-well PCR plate

Sample Volume

15 ~ 120 μL

Maximum Heating and Cooling Rate

≥ 5°C/s

Heating Method

Peltier

Temperature Range

4 ~ 100 °C

Temperature Accuracy (Setpoint)

± 0.1 °C

Temperature Uniformity

± 0.2 °C

Temperature Accuracy (Measurement)

± 0.2 °C

Light Collection Mode

Whole Plate 14s Fast Light Collection Mode

Detector

High Performance PD (Photodiode)

Light Propagation Medium

Optical Fiber

Sample Repeatability

CV ≤ 1%

Linear Range

1 ~ 10¹⁰

Sensitivity

1 copy

Sample Linearity

|r| ≥ 0.999